RELEASE NOTES
Product: FW for MI 3394 (HW1-HW6)
Version: 1.63.41.81220e9f
Release date: 23.11.2023
DESCRIPTION OF THE RELEASED PRODUCT:
FW for MI 3394 CE MultiTesterXA
Description of changes, dependences and history:
MI 3394 |
|
PC SW |
Minimal FW version of related instruments |
|||||
# |
FW version |
Release date |
Description of changes: |
HW |
Metrel User Manual version |
Metrel Electrical Safety Manager version |
A 1460 FW version |
BB protocol Documentation |
28 |
1.63.41.81220e9f |
23.11.2023 |
Changed pre-test (check of relays). |
HW1-HW6 |
HW5 (3.11.23), HW1-HW4 (3.4.13) |
V1.20.1.xxxx |
1.0.0.884 |
1.7.4 |
27 |
1.63.38.55356f5b |
10.10.2023 |
Bug fix, Earth continuity 4-wire adjustment problem (FWD-109) |
HW1-HW6 |
HW5 (3.11.23), HW1-HW4 (3.4.13) |
V1.20.1.xxxx |
1.0.0.884 |
1.7.4 |
26 |
1.63.37.e7546eeb |
04.10.2023 |
Support for i.MX processor Upgraded Discharge Time function, Added visual & functional inspections, Upgraded leakage measurements (added a.c., d.c. components), Added separate Bluetooth initialization, Added option to select worst or last result, Load pretest (disables Riso warning), Support for warning lamps in Single test menu, Support for commander / pedal in Single test menu, Upgraded search function in AutoSequence menu, Updated limit calculator (EN/CSA, NEN3140) IT warning can be disabled during Power on, Added compensation of test leads into AutoSequences, Added R compensation at Continuity measurement, Bug fixes (FW_1.63.XX) |
HW1-HW6 |
HW5 (3.11.23), HW1-HW4 (3.4.13) |
V1.20.1.xxxx |
1.0.0.884 |
1.7.4 |
25 |
1.62.54.10229f4 |
02.06.2023 |
Fixed adjustment procedure for Earth continuity measurement (MPL-1052)
|
HW1-HW6 |
HW5 (3.10.22), HW1-HW4 (3.4.13) |
V1.20.1.xxxx |
1.0.0.884 |
1.7.4 |
24 |
1.62.42.b31bc4f9 |
27.06.2022 |
Improved accuracy @ Continuity 4 wire (added additional range) |
HW1-HW6 |
HW5 (3.9.20), HW1-HW4 (3.4.13) |
V1.18.0.7289 |
1.0.0.884 |
1.7.4 |
23 |
1.62.39.39ffb6f1 |
11.11.2021 |
Support for HW(HD)6, Small bug fixes. |
HW1-HW6 |
HW5 (3.8.19), HW1-HW4 (3.4.13) |
V1.17.0.6962 |
1.0.0.884 |
1.4.3 |
22 |
1.62.38.4d15e5eb |
15.07.2021 |
Support for BT module (ST), Support for new touch screen controller (Goodix), Bug fix, Ethernet MAC address (BMA-484) |
HW1-HW5 |
HW5 (3.8.17), HW1-HW4 (3.4.13) |
V1.17.0.6962 |
1.0.0.884 |
1.4.3 |
21 |
1.62.33.9cd56a1f |
15.02.2021 |
Bug fixes, Leak’s & Power (BMA-473), HV measurement test occasionally not starting (BMA-476), Instrument freezes in certain condition (BMA-474) |
HW1-HW5 |
HW5 (3.8.17), HW1-HW4 (3.4.13) |
V1.17.0.6374 |
1.0.0.884 |
1.4.3 |
20 |
1.62.27.72a1b5f7 |
07.09.2020 |
Bug fix, losing data during Ethernet communication via BB protocol (BMA-470) |
HW1-HW5 |
HW5 (3.7.16), HW1-HW4 (3.4.13) |
V1.16.0.5996 |
1.0.0.884 |
1.4.3 |
19 |
1.62.25.37b37e96 |
14.04.2020 |
Fix of not working flow command, “External OK key mode” flow command (now enabled in the BB protocol) Update of Korean language Bug fix (BMA-423, BMA-460) |
HW1-HW5 |
HW5 (3.7.16), HW1-HW4 (3.4.13) |
V1.15.0.5823 |
1.0.0.884 |
1.4.3 |
18 |
1.62.21.b2b2db0 |
03.12.2019 |
Korean language update Bug fix (HVdc, too low voltage @ certain conditions) |
HW1-HW5 |
HW5 (3.6.15), HW1-HW4 (3.4.13) |
V1.14.0.5457 |
1.0.0.884 |
1.4.3 |
17 |
1.62.9.d4b8e66 |
05.11.2019 |
Support for HW5 version Improved accuracy @ Ipe leakage & Touch leakage current measurements (affects only HW5 or higher versions) Improvements of Riso measuring function (affects adjustment & calibration procedure for HW5 or higher versions) Modified search function in Memory organizer menu Bug fixes (barcode scanner Zebex not working, removed pre-test @ discharge time function, corrected translation of Italian language, various bug fixes) |
HW1-HW5 |
HW5 (3.6.15), HW1-HW4 (3.4.13) |
V1.14.0.5457 |
1.0.0.884 |
1.4.3 |
16 |
1.61.7.11392 |
11.02.2019 |
Upgrade of Black-box protocol (support for execution of single tests) HV AC test voltage raised from 5kV to 5,1kV Integration of custom limits Support for external keyboard, via (A 1578) Move structure element (CutPaste) functionality Enabling Auto save flow command inside memory organizer Support for SmartScandy universal scanner Support for scanners, Socket 7xx Modified method of adding a structure element Modified method of saving after executed AutoSequence QWERTZ, QWERTY, keyboard related to selected language Italian language update |
HW1-HW4 |
3.4.13 |
V1.12.0.5061 |
1.0.0.884 |
1.3 |
15 |
1.59.22.10681 |
24.09.2018 |
Speed up of insulation test (app. 2s) Minor fixes of translations |
HW1-HW4 |
3.3.9 |
V1.9.0.xxxx |
1.0.0.8884 |
1.1.2 |
14 |
1.59.20.10545 |
30.08.2018 |
Fixed error message translation Removed output parameter from Isub function |
HW1-HW4 |
3.3.9 |
V1.9.0.xxxx |
1.0.0.8884 |
1.1.2 |
13 |
1.59.17.9633 |
03.04.2018 |
Support for 127 V supply system Minor bug fixes |
HW1-HW4 |
3.3.9 |
V1.6.0.xxxx |
1.0.0.8010 |
1.1.2 |
12 |
1.59.13.9362 |
26.02.2018 |
Language translation updates (Italian, French) |
HW1-HW4 |
3.3.9 |
V1.6.0.xxxx |
1.0.0.8010 |
1.1.2 |
11 |
1.59.11.9235 |
25.01.2018 |
Bug fix on Bluetooth communication |
HW1-HW4 |
3.3.9 |
V1.6.0.xxxx |
1.0.0.8010 |
1.1.2 |
10 |
1.59.5.9007 |
04.12.2017 |
Supported Able printer AP1310 Various fixes and improvements |
HW2/HW3 |
3.3.9 |
V1.6.0.xxxx |
1.0.0.5136 |
1.1.2 |
9 |
1.58.7.8695 |
20.09.2017 |
Added flow command Auto Print in Autosequences Added new languages: - French - Lithuanian - Swiss-French - Finnish - Portuguese Added Custom Visual inspections in Autosequences Upgraded Auto Save flow command Added QR print tags Supported RFID reader/writer Elatec TWN4 Supported RFID reader Atid AT288 Various fixes and improvements |
HW2/HW3 |
3.2.8 |
V1.5.0.xxxx |
1.0.0.5136 |
1.1.2 |
8 |
1.57.5 |
23.03.2017 |
Fixed bug (affected only on adjustment of the instrument) |
HW2/HW3 |
3.1.7 |
V1.3.0.3555 |
1.0.0.5136 |
1.1.2 |
7 |
1.57.4 |
20.03.2017 |
Power measurement added additional range (HW3), Improved search function, Added user accounts, Support for printer S 2062, Support for scanners A 1321, A 1545, A 1105, Custom limits, Auto tests changed to Auto Sequences, Added comment on measurement, Additional information on the measurement (information about used adapter), Added Serbian language. |
HW2/HW3 |
3.1.6 |
V1.3.0.3555 |
1.0.0.5136 |
1.1.2 |
6 |
1.55.7 |
27.10.2016 |
Added Czech language |
HW2 |
2.4.4 |
V1.3.0.3030 |
1.0.0.5136 |
1.1.2 |
5 |
1.55.4 |
30.06.2016 |
Bug fixed |
HW2 |
2.4.4 |
V1.2.0.2560 |
1.0.0.5136 |
1.1.2 |
4 |
1.55.3 |
16.06.2016 |
Bug fixed |
HW2 |
2.4.4 |
V1.2.0.2560 |
1.0.0.5136 |
1.1.2 |
3 |
1.55.2 |
09.06.2016 |
Support for Black-box protocol enabling remote communication with instrument via RS232, USB, Ethernet, or Bluetooth communication ports. Added Voltage Drop measurement as a sub function of Earth Bond 10A measurement. Added two fields for comment (comment 1, comment 2) under parameters of each measurement. Changed user manual, updated valid “User manual” version is, Ver 2.5.5. |
HW2 |
1.4 |
V1.2.0.2560 |
1.0.0.5136 |
1.1.2 |
2 |
1.54.7 |
29.02.2016 |
Added new multi-language pack: Spanish, Russian, Polish, Hungarian, Slovakian, Danish, Swedish, Norwegian Dutch, Slovenian. Added, new flow commands: Result screen (enabling auto local save), Flow Protocol (enabling remote commands), Appliance info (enabling repeat test, ID increment, appliance type selection). New flow commands are enabling offline complete automatic test with local save option. Added support for barcode – scanner. Added support for A 1460 CE test adapter. Added test lead compensation in continuity 2Wire measurement. Changes in memory organizer (results of auto-test are stored under Node) Complete functionality of the instrument included in basic license |
HW2 |
1.3 |
V1.2.0.2332 |
1.0.0.5136 |
1.1.2 |
1 |
1.53.5 |
11.10.2015 |
Bug Fixes |
HW2 |
1.3 |
V1.1.0.2116 |
|
1.1.2 |
INSTALLATION PROCEDURE:
The instrument can be upgraded from a PC via the RS232 or USB communication port. This enables to keep the instrument up to date even if the standards or regulations change. The firmware upgrade requires internet access and can be carried out from the Metrel ES Manager software with a help of special upgrading software – FlashMe which will guide you through the upgrading procedure. For more information refer to Metrel ES Manager Help file.
COMPANY DATA:
METREL d.d.
Measuring and Regulation Equipment Manufacturer
Ljubljanska 77
SI-1354 Horjul
Tel: + 386 (0)1 75 58 200
Fax: + 386 (0)1 75 49 226
E-mail: metrel@metrel.si
http://www.metrel.si