Instructions
WARNING!!!!
This upgrade includes important upgrades to Auto-tests, which breaks compatibility with the old Auto-test format.
After the upgrade:
- Running old Auto-tests from Auto-tests menu will have strange behaviour.
- Running old Auto-tests from Memory organizer will have strange behaviour (retest or test from a copy or clone of an old Auto-tests result).
- Viewing Auto-test results from Memory organizer will work as expected.
We recommend a quick format of the SD card from the computer to avoid any strange behaviour. Back up any data you want to keep in your PC software.
Before starting the upgrade, make sure that:
- The instrument data is backed-up.
- The upgrading procedure is not interrupted in any way.
As failing to do so, may cause the instrument to malfunction.
Step 1
- Connect the instrument to mains and press power on/off button while holding the enter key to turn the instrument on.
Instrument turned off.
Step 2
- Using an USB or a RS232 cable, connect the instrument to the PC.
USB and RSR232 connectors.
Step 3
- Select the correct COM port. (Hint: for an USB connection the port name is Measurement instrument USB VCOM Port.
Communication port selection.
Step 4
- Click the continue button.
Continue button.